Category: BioWaveHOME

August 23, 2016

Your BiowavePRO display looks dim…

If your BiowavePRO LCD display looks dim or completely blue or completely white, you may just need to adjust the contrast. First, the contrast can only be changed during charging while there normally would be a large battery symbol on the display, or while the unit and electrodes are connected to a patient (so the
June 1, 2016

BiowaveHOME – Tips for plugging in and unplugging the cable

If the BioWaveHOME cable needs to be removed, FIRST place the BioWaveHOME unit FACE DOWN on a hard flat surface. Next, use your index finger on one hand to firmly press the nipple on the cable release button while the other hand GENTLY pulls out the connector. Having said this, the cable NEVER has to
May 31, 2016

BiowaveHOME – Unit Throttles Back Intensity from 15% to 0%

This means the skin impedance AND/OR the electrode impedance is too high. The software that controls BiowaveHOME works differently than in BiowavePRO. If a patient’s skin is too dry (impedance is too high) or if the noninvasive electrode (pad) has been used many times and is dry or not tacky to the touch (impedance is